Chapter 6 - Service Adjustments
6-5
6-1-2-5-1
Adjusting the Blue Tint
NOTE:
The default blue tint test pattern is set at two bars - see
1.) Press
Alt Q
to access the Display Calibration Screen.
2.) To adjust the blue tint: hold down <ALT>, then use the < and > keys, as required.
Repeatedly pressing < will decrease the blue tint; repeatedly pressing > will increase the blue tint.
NOTE:
To assist in making blue tint adjustments, an adjustment bar is displayed below the screen -
see
.
3.) Press
Alt Q
to exit to the Display Calibration Screen.
Figure 6-7 LCD Display Calibration Screen showing Test Pattern
Figure 6-8 Blue Tint Adjustment Bar
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...