GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 4 - General Procedures and Functional Checks
4-35
NOTE:
The scale value may affect FPS, Low Velocity Reject, and Sample Volume.
•
Adjust
Low Velocity Reject
to remove low velocity blood flow and tissue movement that reduces
image quality.
•
Adjust
Variance
to detect flow disturbances.
•
Adjust
Sample volume
(SV) to a low setting for better flow resolution, or a higher setting to more
easily locate disturbed flows
•
Adjust
Frequency
to optimize the color flow display. Higher settings improve resolution. Lower
settings improve depth penetration and sensitivity. This does not affect the frequency used for 2D
and M-Mode.
NOTE:
The Frequency setting may affect FPS, SV and Low Velocity Reject.
•
Adjust
Power
to obtain an acceptable image using the lowest setting possible.
NOTE:
The Power setting affects all other operating modes.
Adjust the following settings to further optimize display of the image:
•
Use
Invert
to reverse the color assignments in the color flow area of the display.
•
Use
Tissue priority
to emphasize either the color flow overlay, or the underlying grey scale tissue
detail.
•
Use
Baseline
to emphasize flow either toward or away from the probe.
•
Use
Radial
and
Lateral Averaging
to reduce noise in the color flow area.
Radial and Lateral Averaging smooths the image by averaging collected data along the same
horizontal line. An increase of the lateral averaging will reduce noise, but this will also reduce the
lateral resolution.
4-4-3-5
Select Color M Mode
1.) Select M Mode (see:
).
2.) Use the trackball (assigned function: Pos) to position the ROI frame over the area to be examined.
3.) Press
Select
. The instruction
Size
should be highlighted in the trackball status bar.
NOTE:
If the trackball control Pointer is selected, press trackball to be able to select between Position
and Size controls.
4.) Use the trackball to adjust the dimension of the ROI.
4-4-3-6
Adjust the Color M Mode Controls
•
Adjust the
Active mode gain
to set the gain in the color flow area.
•
Adjust
Scale
to the highest setting that provides adequate flow detection.
NOTE:
The scale value may affect FPS, Low Velocity Reject, and Sample Volume.
•
Adjust
Low Velocity Reject
to remove low velocity blood flow and tissue movement that reduces
image quality.
•
Adjust
Variance
to detect flow disturbances.
•
Adjust
Sample volume
(SV) to a low setting for better flow resolution, or a higher setting to more
easily locate disturbed flows.
•
Adjust
Frequency
to optimize the color flow display. Higher settings improve resolution. Lower
settings improve depth penetration and sensitivity. This does not affect the frequency used for 2D
and M-Mode.
NOTE:
NOTE: Frequency setting may affect FPS, SV and Low Velocity Reject.
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...