6-14
-
6.) Leave the system overnight to complete the Full Discharge process.
The Vivid q N system automatically shuts down at the end of this process; the battery calibration
procedure is now complete.
7.) The following morning, re-connect the AC power cable and turn ON power to the system.
NOTE:
The system may now be used normally, without any dependence on battery capacity status.
Figure 6-14 Battery Calibration Wizard - Page 3