Section 10-6 - Electrical Safety Tests
10-6-8-1
Data Sheet for ECG Current Leakage
The test passes when all readings measure less than the values shown in the tables below. Record all
data.
Table 10-17 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
AC Power
Source
Maximum
Allowance Limit
GROUND
OPEN
GROUND
CLOSED
Patient Lead-to-Ground Current Leakage Test
and
Patient Lead-to-Lead Current Leakage Test
115V
10uA
10uA
220/240V
500uA
10uA
Table 10-18 Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power
Source
Maximum
Allowance Limit
Patient Lead Isolation Current Test
115V
20uA
220/240V
5mA
Table 10-19 Typical Data Sheet for ECG Lead-to-Lead Test
ECG
Power
Tester
Polarity
Switch
Tester
Ground
Switch
Tester Lead Selector
RL
RA
LA
LL
C
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...