Chapter 1 - Introduction
1-11
1-4-4
Electrical Safety
To minimize shock hazard, the equipment must be connected to a well grounded power source. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety grounding.
To ensure proper grounding, connect this equipment to a receptacle marked “HOSPITAL ONLY” OR
“HOSPITAL GRADE”.
The power outlet used for this equipment should not be shared with other types of equipment. Both the
system power cable and the power connector must meet international electrical standards.
1-4-4-1
Probes
All the probes for the Vivid q N ultrasound unit are designed and manufactured to provide trouble-free,
reliable service. To ensure this, correct handling of probes is important and the following points should
be noted:
•
Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
•
Do not use a cracked or damaged probe. In this event, call your field service representative
immediately to obtain a replacement.
•
Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
•
To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the SafeLock Cart wheels.
Follow these guidelines before connecting a probe to the scanner:
•
Inspect the probe prior to each use for damage or degradation to the:
-
housing
-
cable strain relief
-
lens
-
seal
-
connector pins
-
locking mechanism
•
Do not use a damaged or defective probe.
•
Never immerse the probe connector or adapter into any liquid.
•
The system has more than one type of probe port. Use the appropriate probe port designed for the
probe you are connecting.
NOTE:
For detailed information on handling endocavity probes and invasive intracardiac probes, refer
to the appropriate supplementary instructions for each probe. In addition, refer to the Vivid q N
User Manual for detailed probe handling instructions.
WARNING
WARNING
CONNECTING A VIVID Q N SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY IT.
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...