Chapter 6 - Service Adjustments
6-13
5.) Disconnect the AC power cable from the Vivid q N, then click
Next
.
The second phase of the process (Full Discharge) commences and the Battery Calibration Wizard
(Page 3) is displayed as shown in
Figure 6-13 Battery Calibration Wizard - Page 3