GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
4-4
Section 4-2 - General Procedures
NOTE:
If the system cannot boot up, or it boots up with a PS communication error, refer to the instructions for
8-3-4 "BIOS Battery Replacement" on page 8-47
.
Follow These Steps to Connect Mains Power to the Unit
Connecting the Vivid q N ultrasound unit involves preliminary checks of the power cord, voltage level
and compliance with electrical safety requirements.
1) Ensure that the wall outlet is of appropriate type, and that the power switch is turned OFF.
2) Uncoil the power cable, allowing sufficient slack so that the unit can be moved slightly.
3) Verify that the power cable is without any visible scratches or any sign of damage.
4) Verify that the on-site mains voltage is within the limits indicated on the rating label on the base of
the Vivid q N portable ultrasound scanner.
5.) Connect the Mains Power Cable female plug via the AC Adapter to the Power Inlet (DC IN - refer
) at the rear of the unit.
6.) Connect the other end (male plug) of the Mains Power Cable to a hospital-grade mains power outlet
with the proper rated voltage.
The unit is now ready for Power ON/Boot Up.
.
Figure 4-1 Mains Power Connector - Rear Panel
DC IN
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...