Section 7-4 - Performing Front End (FE) Diagnostics
7-4-2-2
DC Offset Verification
NOTE:
For a description of this test, refer to the information provided in
"DC Offset Verification" on page
.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the
Diagnostic
Test
window, trackball to
DC Offset Verification
in the list of tests available for
selection from the Diagnostic Test Tree (see
) and press
Set
. The DC Offset
Verification name is displayed under
Execute;
the relevant system components to be tested are
highlighted light blue in the Schematic View (PM, TR32 and RFI boards), as shown below:
2.) As indicated in the
Special Instructions
area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the
Special Instructions
area, activate the
Loops
checkbox
and select the required number of
Loops
from the combo box.
4.) Adjacent to
On Fail,
select either
Stop
or
Continue
, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the
Start
button and press
Set
.
The DC Offset Verification Test commences. While the test proceeds, the name of the test currently
in progress (DC Offset Verification) is displayed in the
Execute
field and the current
Loop #
and
specific test are shown below. The messages
Init Done
and
Please
Wait
are displayed beside the
progress bar in the
Status
area.
Figure 7-9 DC Offset Verification Test
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...