GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
5-44
Section 5-9 - ECG/Respiratory Module
5-9-2-2
Specifications
5-9-2-3
Connectors
Table 5-17 External ECG/Respiratory Interface Specifications
Characteristic
Specification
Input Range (ac)
2v peak to peak
A2D resolution
12 bit ,2'complement
Sample Rate
500 [samples/second]
Ground Isolation
4 kV rms
Table 5-18 Connector Pin Out J5 (ECG, Respiration, External ECG)
Pin Number
Specification
1
External ECG Input
2
ECG Input, Respiration Input/Output (LL)
3
ECG Input, Respiration Input/Output (RA)
4
ECG Respiration Output (LA)
5
External ECG Select Signal Input (Short to GND on Cable)
6
GND (GND for External ECG or Shield for ECG)
1
3
2
4
5
6
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...