Chapter 1 - Introduction
1-13
•
Patient Environment IEC 60601-1-1
Sub clause 2.204
Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out.
It is very difficult to attach unique dimensions to the PATIENT ENVIRONMENT. In practice a
distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel stand and a horizontal
distance of 1,5 m (4.9 ft.) have justified themselves as indicative of the dimensions of the Patient
Environment. The patient environment/vicinity is depicted as a dashed line in this procedure - see
the example in
.
•
Patient Environment EN 60601-1
Sub clause 3.79 - Patient Environment
It is difficult for this standard to define dimensions for the volume in which diagnosis, monitoring or
treatment occurs. The dimensions for the PATIENT ENVIRONMENT given in
have been
justified in practice.
Figure 1-1 Patient Safety Environment
Figure 1-2 Example of Patient Environment
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...