Chapter 8 - Replacement Procedures
8-13
8-2-4-5-1
Keycap Removal Procedure
NOTE:
Replacement Keycaps for the Vivid q N system are supplied as part of the Keycap Kit - Part No.
S2423359. (For Scandinavian countries, Part No. 2420372-1).
1) Carefully place a small flat screw driver between the keycap you wish to remove and its neighboring
keycap.
2.) Use the screw driver to gently lever off the defective keycap, until it is completely loosened from its
).
3.) Lift off and remove the defective keycap.
8-2-4-5-2
Keycap Installation Procedure
1) Carefully place the appropriate new keycap in position on the keyboard, taking care to place the
plastic alignment pin in the correct position so that the keycap is the right way up and reads
correctly.
2.) Gently push the keycap down until it snaps into position.
Figure 8-17 Keycap Replacement
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...