GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-49
5.) Carefully lift the DC power supply
upwards
and remove it from the Bottom Assembly, as shown in
6.) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the Vivid q N system is
currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
-
"Calibration Tests" on page 7-25
-
"Full System Test" on page 7-10
Figure 8-57 Loosening the Screws on the DC Power Supply
Figure 8-58 Removing the DC Power Supply