Chapter 4 - General Procedures and Functional Checks
4-25
4-3-2-4
Wireless USB Network Test
NOTE:
When the Wireless Network USB Adaptor is installed in a new system, the functionality is verified as
described in
"Configuring the Wireless Adaptor" on page 3-93
.
NOTE:
When the Wireless Network USB Adaptor is installed in a system that is already configured, proceed as
described below.
4-3-2-4-1
Wireless USB Network Functionality Test
1.) With the system up and running, press
Config
.
2.) Click
Connectivity
and select
Dataflow
- see
.
3.) From the Dataflow
Name
field scroll-down list (
), select a dataflow that the user works
with in day-to-day operations.
4.) Under
Selected devices
), select the appropriate device.
5.) Click
Check
.
The Information - Check Status box opens -
Figure 4-7 Dataflow showing “Remote Database” as the Selected Device
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...