Section 7-4 - Performing Front End (FE) Diagnostics
NOTE:
For this test, there are two reports (MLA0 and MLA1) which is listed in Reports on the right. By
default, the MLA0 report is displayed in the Reports window below, showing the results for both
the TR32A and TR32B boards (
9.) To view the MLA1 test results in the Report window, select MLA1 test results under
Reports
.
The MLA1 report is displayed in the Reports window below:
10) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
11) After reviewing the test results, proceed as follows:
•
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
•
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
•
If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in
"DC Offset Calibration" on page 7-25
, and then repeat the test.
•
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in
Components and Function (Theory)
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
12.)Trackball to the
Exit
button and press
Set
to close the
View Test Log
dialog box.
13.)Press the <
Esc
> button on the keyboard to close the Diagnostic
Test Window
.
Figure 7-23 View Test Log - TR32 BF Digital Test (MLA1)
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...