Chapter 8 - Replacement Procedures
8-61
7.) Disconnect the LCD flex
[1]
from the rear panel of the LCD Display and remove -
8-3-9-5
AUO LCD Cable and Flex Installation Procedure
NOTE:
The numbers shown in square brackets correspond to the numbers listed in
1.) Connect the LCD Flex to the LCD as shown in
2.) Thread the Flex Board cable through the Display Flex Cover
[6]
making sure that the Display Flex
Cover is correctly orientated as shown in
.
3.) Thread the Flex Board cable through the ferrite core
[7]
and secure the ferrite with a strip of Kapton
tape -
4.) Connect the Inverter Flex Board cable
[2]
to the Flex Board
[4],
then connect its other end to the
LCD Backlight Inverter board
[5] -
.
5.) Connect the LCD Backlight Inverter board
[5]
to the AUO LCD cable
[3] -
.
Figure 8-71 LCD Flex Connected to LCD Rear Panel
Figure 8-72 Flex Board Cable Threaded through Flex Board Cover and Ferrite
6
7
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...