Chapter 8 - Replacement Procedures
8-11
remove it from the Vivid q N and place it next to the system, as shown in
.
NOTE:
When removing the Control Panel and Keyboard during other replacement procedures, Step 6
will only be required if indicated.
6.) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the control panel/keyboard cable from the system as shown in
Figure 8-15 Removing the Control Panel and Keyboard
Figure 8-16 Disconnecting the Control Panel/Keyboard Cable
Disconnect here