Section 8-2 - External Component Replacement Procedures
Section 8-2
External Component Replacement Procedures
8-2-1
Bearing Handle Replacement Procedure
NOTE:
The term “Bearing Handle” relates to the Vivid q N portable carrying handle.
8-2-1-1
Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Bearing Handle replacement
procedure.
8-2-1-2
Time Required
10 minutes
8-2-1-3
Preparations
None
8-2-1-4
Bearing Handle Removal Procedure
1) Close the Vivid q N LCD cover, making sure both front latches are securely fastened (
).
2.) Release the captive screw on either side of the Bearing Handle (there are two screws that secure
the handle in position), by turning 1/4 turn in a
counterclockwise
direction (
).
Figure 8-1 Closing Vivid q N LCD Cover
Figure 8-2 Releasing the Bearing Handle Captive Screw
Latch
Latch
Captive Screw
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...