8-44
Section 8-3 - Internal Component Replacement Procedures
8-3-3-5
BEP Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the
NOTE:
The
SATA BEP
has an adaptor bracket, which is required for its installation. The bracket is
shown on the left in
; in the picture on the right, the bracket is shown in position in
the Vivid q N scanner. Notice that the bracket sits on the PCMCIA slot-flap, which is pressed
open while the adaptor bracket is positioned on the flap itself. To install the adaptor bracket, the
flap is pressed open (
) and the adaptor bracket is placed in position. The procedure
is described below.
Figure 8-52 SATA Adaptor Bracket
Figure 8-53 PCMCIA Slot Flap Pressed Open
SATA Adaptor Bracket
SATA Adaptor Bracket
PCMCIA Slot Flap
PCMCIA Slot Flap Ready for Positioning the Adaptor Bracket
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...