Chapter 8 - Replacement Procedures
8-45
8-3-3-5-1
Preparation Procedure
1) Press the PCMCIA slot-flap open and hold it down as shown in
2.) Place the adaptor bracket in position as shown on the right in figure
and press it firmly
against the Vivid q N chassis.
NOTE:
shows
a view from the rear of the
Vivid q N
showing the adaptor bracket closing off
the PCMCIA slot.
3.) Connect the control panel/keyboard cable (if previously removed from an ATA BEP; see
) to the SATA BEP.
Proceed to install the SATA BEP as described in the next procedure.
8-3-3-5-2
BEP Installation
1) Carefully return the new BEP to the original position on the Bottom Assembly - refer to
NOTE:
When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access
codes already registered in the Hard Disk. Only qualified personnel should perform these tasks.
2.) Check that the BEP is seated correctly then return and fasten the four screws (previously removed)
to secure the BEP to the Bottom Assembly - refer to
3.) Re-connect all the cables to the BEP:
Proceed as follows (refer to
):
a.) Re-connect the Speaker-to-BEP power cable
b.) Re-connect the Fan-to-BEP power cable.
c.) Re-connect the HVPS-to-BEP power cable.
d.) Re-connect the BEP-to-RFI Flex cable.
e.) Re-connect the LCD Flex ribbon cable (carefully push back and secure the two connector
latches).
4.) Return and fasten the two securing nuts previously removed (one on each side of the VGA
Connector, located on the Vivid q N rear panel), as shown
.
5.) Return the Control Panel and Keyboard, as described in the
Installation Procedure" on page 8-12
6.) Return the Hard Disk, as described in the
"Hard Disk Installation Procedure" on page 8-7
.
Figure 8-54 Rear View of the Vivid q N Scanner after SATA BEP Installation
VGA Connector
Adaptor Bracket
(external view)
Hard Disk shown without
its retaining screws
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...