Chapter 8 - Replacement Procedures
8-73
8-4-6
Lower Cover (Rear) Replacement Procedure
8-4-6-1
Tools
Appropriate Phillips screwdriver.
8-4-6-2
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-4-6-3
Lower Cover (Rear) Removal Procedure
Remove six retaining screws of the lower cover (rear) and while supporting the cover with one hand,
gently remove the cover - see
8-4-6-4
Lower Cover (Rear) Installation Procedure
1) Return the cover to the correct position on the rear of the SafeLock Cart - see
2.) While holding the cover to prevent it from falling, fasten the retaining screws of the lower cover
(rear).
3) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-86 Lower Cover (Rear) Removal
Retaining Screw
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...