Chapter 4 - General Procedures and Functional Checks
4-23
4-3-1-6
CW Test
4-3-1-7
Visual Test
Table 4-10 CW Test - Recommended
Step
Task
Expected Result(s)
1
Boot-up the System in Regular Scanning Mode
2
Press
Alt + Config
to enter Diagnostics Mode
3
From the Diagnostic Test tree, select
Front End Test
.
4
Run
Full System Test
and press
Start
PASS
5
Connect the 3S-RS probe to the probe port.
6
From the Diagnostic Test tree, select
Front End Tes
t.
7
Select
TR32
then select
TX
. Select
TXProbeTX Test
and press
Start
PASS
8
Select
MUX Test
then select
3S-RS Test
and press
Start
PASS
9
Select
TR32 Test
then select
RX Tes
t. Select
RSTest
with 3S-RS
and press
Start
PASS
10
When done, press the
Esc
button
Table 4-11 Visual Test - Recommended
Step
Task
Expected Result(s)
1
Insert 3S-RS probe.
2
Increase the gain to see the homogenous noise.
The probe ID is 3S-RS. No strikes or artefacts are
present.
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...