Chapter 8 - Replacement Procedures
8-33
8-3-2-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-3-2-2
Time Required
15 minutes
8-3-2-3
Preparations
Shut down the Vivid q N ultrasound unit, as described in
CAUTION
When removing the boards, ESD may cause damage to a board. Always have the ESD hand strip
connected to the machine chassis and to your hand.