Chapter 7 - Diagnostics/Troubleshooting
7-61
7-4-5
Probe and MUX Diagnostic Tests
7-4-5-1
Accessing the Probe and MUX Diagnostic Test Options
•
Access the
Diagnostic
Test
window, as described in the
Accessing the Front End Diagnostic
7-4-5-2
ACW Test
7-4-5-2-1
ACW IQ Symmetry Test
NOTE:
For a description of this test, refer to the information in
"ACW IQ Symmetry Test" on page 7-13
.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the
Diagnostic
Test
window, trackball to
ACW IQ Symmetry Test
in the list of
ACW
tests
available
for selection from the Diagnostic Test Tree (see
) and press
Set
. The
ACW IQ Symmetry Test name is displayed under
Execute;
the relevant system components to be
tested are highlighted light blue in the Schematic View (P&M), as shown below:
2.) As indicated in the
Special Instructions
area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
It is necessary to
disable
the Loop option where it should not be used.
After performing this test, you must re-boot the system and perform the DC Offset Calibration procedure
as described in
"DC Offset Calibration" on page 7-25
.
Figure 7-35 ACW IQ Symmetry Test
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...