GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
10-12
Section 10-4 - System Maintenance
10-4-4-2
Disinfection
The disinfection procedures described in this section should be routinely performed by medical
personnel, as shown in
. The information is provided in this manual for reference purposes.
NOTICE
IMPORTANT - To avoid contamination when the system is in use during operations, always
cover the keyboard and the monitor with the OR covers provided for this purpose.
Table 10-8
Routine Disinfection Procedures
What?
When?
How?
Materials?
Who?
Keyboard and Trackball
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Switch the system off!
Disinfectant with a
70:30 mixture of
isopropyl alcohol and
water.
Never allow liquid to
flow into the keyboard!
Use a 70:30
mixture of
isopropyl
alcohol and
water.
Staff users/doctors
Monitor display-surface
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Clean with dry soft,
lint-free cloth.
When contaminated
with blood or other
stains secretions, use a
70:30 mixture of
isopropyl alcohol and
water.
Soft,
lint-free cloth.
Disinfect with a
70:30 mixture of
isopropyl
alcohol and
water.
Staff users/doctors
Devices' external surfaces
(incl. monitor housing, handle
and SafeLock Cart wheels)
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Clean the washable
outer surfaces with a
soft, non-abrasive cloth
including the SafeLock
Cart wheels. Do not use
abrasive powder or
strong
chemicals!
Use a soft,
non-abrasive
folded cloth or
sponge with a
mild, general
purpose,
non-abrasive
soap and water
solution.
Trained cleaning staff
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...