Chapter 3 - System Setup
3-59
5) Align the connector with the probe port and carefully push into place, as shown in
6) Press the connector locking lever
upwards
to the full vertical position to lock in place, as shown in
.
7) Carefully position the probe cord so that it is free to move and is not resting on the floor.
When the probe is connected, it is automatically activated.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
Figure 3-58 Connecting Probe to the Vivid q N
Figure 3-59 Securing Probe with Locking Lever
CAUTION
Do not bend the probe cable acutely
.
Fault conditions can result in electric shock hazard.
Do not touch the surface of probe connectors which
are exposed when the probe is removed. Do not touch the patient when connecting or disconnecting a
probe.
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...