Section 10-6 - Electrical Safety Tests
•
Check probe current leakage, as shown below:
NOTE:
Because the Vivid q N scanner is operated from low voltage DC (considered extremely safe
extra low voltage) and because the transformer is enclosed in plastic parts only (and therefore
does not have any accessible metal parts),
when the scanner is not mounted on the
SafeLock Cart
, it is not necessary to perform the following safety tests (i.e., they apply only to
the SafeLock Cart itself):
•
Chassis Current Leakage Test
•
Grounding Continuity Test
Probe
Frequency for Checking
Surface
Once per year
(or in accordance with QA program)
Endocavitary
Twice per year
(or in accordance with QA program)
Any probe
Whenever damage is suspected
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...