GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 188
Section 8-7 - Main Console parts replacement
2.
Flip the Fan Tray over on a clean, safe
surface.
Pull out the Filter latch and slide the Filter
out.
Clean the Bottom Filter after it is removed
by removing excess lint or dust from the
soiled side. If necessary, use a clean, soft
brush; or vacuum.
3.
Reinstall the Filter into the Fan Tray and reinstall the Fan Tray.
Make sure the Fan Tray seats
completely into the Card Cage.
Table 8-114 Removing the Fan Tray Filter