GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 3 LOGIQ E9 Setup
3 - 63
3-9-4-6
How to Setup and Use a DICOM MPPS (Modality Performed Procedure Step) Service
A DICOM Performed Procedure, also known as Modality Performed Procedure Step (MPPS), is a way
to inform your billing system that an ultrasound procedure has been completed and can be billed.
A typical setup has your Worklist Service handle MPPS communications. For example, you can retrieve
today’s scheduled exams and report completed exams to the same system.
To set up MPPS on the , follow these steps:
Table 3-22 Setup and Use a DICOM MPPS
Utility
->
Connectivity
->
Service
1.
In
Utility
->
Connectivity
->
Service
, from the Destination Device drop-down menu, select
your Worklist device.
2.
From the drop-down menu of services, select
Dicom Performed Procedure
and press
Add
.
3.
Provide a name, set the AE Title and Port Number for your Worklist service. AE Title is case-
sensitive.
4.
Maximum Retries, Retry Interval and Timeout have the same use as with DICOM Image
Storage Services. See:
3-9-4-2 "Properties" on page 3-45
.
5.
Save your changes and press
Verify
. A Smiley Face indicates successful communication with
the service.
(continued on next page)
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......