4 - 36
Section 4-3 - Functional checks
4-3-7
System CFM and PWD Checks
4-3-7-1
Introduction
For information on the system’s different modes as well as expected results, the Basic User Manual or
User Guide will familiarize you with image optimization for
B-Mode
,
M-Mode
,
Color Flow
, and
Doppler
.
For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter
5.
4-3-7-2
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
-
see:
Section 3-6-6 "Connecting Probes" on page 3-20
for info about connecting the probes.
-
For available probes, see:
Section 9-18 "Probes" on page 9-104
.
2.) Turn ON the scanner.
4-3-7-3
Color Mode Checks
Color Flow screens are B or M-Mode screens with colors representing blood or tissue movement. Color
Flow may be selected both from B-Mode, or from M mode, or a combination of these.
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-4
Color B-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-5
Color M-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
•
Check Horizontal Sweep, PRF, Baseline, and Invert
•
Check Variance, Color Maps and Cineloop
•
Check Sample Volume
•
Check Frequency and Acoustic Power
•
Adjust ROI
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......