GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 23
7-5-3 Diagnostics Window Overview (cont’d)
Figure 7-24 Diagnostic Window Overview
The Service Diagnostics windows have some advanced features.
NOTE:
The Loop Count field must have a value of 2 or greater to activate the Advanced features.
8.) This button toggles between Standard and Advanced mode. The window below shows the
advanced features.
9.) Select the choice for Stop On Failure for the Looping and List sections:
-
Yes = the test will abort at the first failure
-
No = the test will complete even if there are failures
10.)Select the choice for Stop On Abort for the Looping and List sections:
-
List = the list will abort when you click the Abort button
-
Loop = the looping will abort when you click the Abort button
-
Diag = the diagnostic test will abort when you click the Abort button
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......