GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 293
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet and all Probes.
3. Allow the Capacitor Pack to discharge. LEDs OFF.
4. Remove the Left Side cover, the Top Cover and WLAN Antennas, if present.
5. Remove the Capacitor Pack.
Preparation Links (if you need more information):
• 4-2-3 "Power shut down" on page 4-7.
• 8-5-2 "Side Covers replacement" on page 8-30.
• 8-5-4 "Top Cover replacement" on page 8-38.
Table 8-183 Preparations and Preparation Links