GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 85
7-5-12-1 Keyboard (cont’d)
Footswitch Tests
Under Service Diagnostics / BEP Interactive/ Keyboard
INPUT > Press left, middle or right footswitch keys.
OUTPUT > Keys will be registered as follows:
-
Left Footswitch = Left W F10
-
Middle Footswitch = Left W F11
-
Right Footswitch = Left W F12
7-5-12-2
Mouse (Trackball)
See:
7-5-3 "Diagnostics Window Overview" on page 7-22
for a description of this window’s features.
Runs a test on essential functions of the Trackball. The test results are displayed in the Status portion
of the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Verify the signal cable is connected securely.
4.) Clean the Trackball. If the problem continues:
5.) Replace the Trackball.