GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 3 LOGIQ E9 Setup
3 - 49
3-9-4-3
Sending Images and Data to a DICOM Image Storage Service
The provides three ways to send images and data:
1.) Dataflow
2.) Print Button
3.) Send To
1.) Dataflow
The default setup stores images in its local Hard Disk Drive. A dataflow allows the to send images to
other locations as well.
To setup a dataflow, follow these steps:
Table 3-12 Dataflow
Utility
->
Connectivity
->
Dataflow screen
1.
Navigate to
Utility
->
Connectivity
->
Dataflow
; press
Add
. The system creates a dataflow
called “NewDataflow.”
2.
Change the name to one of your choosing.
3.
Notice the list of available services to the left of the
<<
and
>>
buttons. Select each DICOM
service you want to attach to your dataflow and press
>>
. This makes it appear in the list on
the right, indicating that the service is now assigned to the dataflow.
NOTE: You cannot remove “Local Archive – Int HD” from the dataflow. This guarantees that
images will be saved on the local Hard Disk Drive of the and not just transmitted to the service.
4.
Check
Direct Store
if you want an image generated by a print button to immediately be sent
to the dataflow’s attached services rather than waiting until the exam is ended for all images.
NOTE: If you plan to generate long cine loops, “Direct Store” is NOT recommended (either on
the dataflow or via a Print button directly), because system response may be slow.
(continued on next page)
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......