GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 22
Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-6-8
Verifying the software option key settings
NOTE:
Software Option Dongles are software version dependant. If a new software version has been
installed, follow installation instructions for that software. Usually, new software option strings
are needed. When reinstalling the same software, the Software Option Strings should be
reused.
1.) From the Touch Panel, select
Utility
->
Admin
->
System Admin
.
2.) Verify that the Software Option String(s) (alphanumeric strings) in the S/W Option Dongle section
is (are) the same as recorded in
Table 4-10 "Software Option Keys" on page 4-26
.
3.) If needed, enter any new or missing option key string(s) in S/W Option Key field and select
Add
.
8-4-6-9
Verifying the TCP/IP settings
1.) From the Touch Panel, select
Utilities
->
Connectivity
->
TCP/IP
.
2.) Verify that all settings on the TCP/IP screen match those recorded in
Table 4-2 "Record settings
from TCP/IP screen" on page 4-18
.
3.) Select “
SAVE SETTINGS
.“
4.) Select
OK
to continue.
5.) Reboot system.
8-4-6-10
Verifying the AE Title and Port settings
1.) From the Touch Panel, select
Utilities
->
Connectivity
->
Device
.
2.) Verify that all settings on the Device settings screen match those recorded in
4-2-11-4 "Connectivity
— Recording the AE Title and Port settings" on page 4-20
.
3.) Select “
SAVE SETTINGS
.”
4.) Select
OK
to continue.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......