Chapter 8 Replacement procedures
8 - 189
8-7-4-1
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-7-4-2
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required checks and is ready for use.
Table 8-115 Rear Filter Installation Functional Checks
See:
Section
Functional Check
Debrief Script
4-2-2
Power ON/Boot Up
LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required
checks and is ready for use.
4-2-3
Power shut down
4-3-14
Probe/Connectors Checks
4-3-5
B-Mode Checks
4-3-7
System CFM and PWD Checks
4-3-18
Mechanical Function Checks