Chapter 5 Components and Functions (Theory)
5 - 11
5-2-7 Signal flow overview (cont’d)
The MRX Board combines the functionality of the Receiver Boards (DRX and GRX) and the GFI.Weak
ultrasound echoes from body structures and blood cells are received by the probes and routed via the
Relay board and the XD bus to the MRX board. The MRX board amplifies the ultrasound signal and
connects it with an A/D converter to the digital domain. The digital signals are then further processed
on the MRX boards.
Figure 5-6 LOGIQ E9 Signal Flow
- MRX
GTX 64 Ch
GTX
64 Ch
GTX 64 Ch
256
256
MRX
Frontplane
GRLY
256
R
e
l
a
y
s
192
192
192
GDIF
David
1
David
2
Pulser
David
3
Pulser
...
David
0
T/ R
A
/
D
A/DA
Nathan
32 Ch
TGC
TGC
TGC
TGC
TGC
X
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
MRX 192 Ch
X
I
CW
Option
Analog
Beamformer
Filter
A/D
Converters
I / Q
Front End
Interface
GFE
SED
DSP
Test
Gen
TGC
Gen
X
A/DA
A/DA
A/DA
Q
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......