GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 6 Service Adjustments
6 - 9
Section 6-4
Operator Panel movement
6-4-1
Purpose of this section
The Operator Panel movement within the LOGIQ E9 is described in this section.
6-4-2
Adjusting the XY Locking Mechanism
Table 6-4
Adjusting the XY Locking Mechanism
Steps
Corresponding Graphic
1.
This procedure is intended to adjust the
locking mechanism for the XY Assembly.
Locate hex screw behind the XY Assembly
to adjust it correctly.
Use a 3 mm hex tool to adjust. Rotating the
screw clockwise will tighten and rotating
the screw counterclockwise will loosen the
lock.
Too tight will reduce the slack and make it
harder to engage the lock. Make the
adjustments in 1/4 turn increments and test
the lock until the optimum adjustment is
achieved.
NOTE: When turning the hex screw, do not
exceed half turn increments.
Overtightening will prevent the console
from locking into place and too loose will
make the console loose. You may try 1/2
turn adjustments, but the console may be
too tight or have too much slack.
XY Adjustment for Operator Panel