8-11-9-1
Shear Wave Capacitor Pack removal
NOTE:
The Shear Wave Capacitor Pack is removed and attaches to the BEP in the same manner as
the Printer Tray.
Table 8-184 Capacitor Pack removal
Steps
Corresponding Graphic
1.
Raise the console height to the highest level.
2.
If the LOGIQ E9 has WLAN, remove the
Antennas on the top of the BEP.
3.
Disconnect the power cable from the Main
Power Supply to the Capacitor Pack
BCM1