GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 3
7-3-3
Collecting a Screen Capture with Logs
If the system malfunctions, press the Alt+D keys simultaneously. This Alt+D function is available at all
times, and collects a screen capture of the image monitor, user-defined presets, and the following logs:
•
Keyboard Shadow Log
•
Error Logs
•
Crash Log
•
Vital Product Data
•
DICOM Logs
•
Windows Event Logs
•
Diagnostic Logs
•
Service Logs
For a detailed list of Service Logs captured, see:
7-3-4 "Capturing Service Logs with ALT+D" on page
7-5
.
When Alt+D is pressed, a menu box opens. Enter the following information:
•
System ID serial number.
•
Software version.
•
System Date and time of occurrence.
•
Sequence of events leading to issue.
•
Is the issue repeatable?
•
Imaging mode, probe, preset/application.
•
Media brand, speed, capacity, type.
•
Select the Destination (storage media or
Service
directory for remote viewing through InSite ExC)
and click the Store button.
Figure 7-1 ALT+D Dialog Box
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......