GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 258
Section 8-10 - Main Power Supply parts replacement
8-10-4-4-2
4D MC Replacement for Cherokee/Mitra PS
NOTE:
Discard any removed components in the appropriate manner.
Table 8-165 4D MC Removal for Cherokee/Mitra PS
Steps
Corresponding Graphic
1.
Locate the 4D Motor Controller on the
back of the Main PS.
Remove the five screws (circled) securing
the 4D MC Cover to the Main PS.
Slide cover off to access the
4D MC.
4D MC Cover removed
2.
Disconnect the USB Cable (
1
) and the
Flex Cable (
2
) from the PS. DO NOT pull
from the ribbon to remove. Only pull from
the connector.
Remove two screws (circled) and slide the
board away to remove the 4D Motor
Controller.
NOTE: The replacement 4D MC includes
a new Flex Cable, two USB Cables and a
P-clamp.