Section 6-3 - DC Offset Calibration
Section 6-3
DC Offset Calibration
6-3-1
Introduction
The DC offset calibration is performed to calibrate each ADC channel of the GRX inputs. The system
calibrates the ADC to zero output when there is no signal on the input. These bias voltages are stored
in the Back End Processor.
6-3-2
When to do a DC Offset Calibration
Do DC Offset Calibration:
•
when installing a new LOGIQ E9
•
after software has been updated or replaced
•
when DRX board(s) or a MRX board has been interchanged
•
when a DRX board(s) or a MRX board has been replaced
•
when the Back End Processor has been replaced
6-3-2-1
DC Calibration Procedure
1.) Disconnect all connected probes.
2.) Power on the LOGIQ E9.
3.) On the Touch Panel, select
Utility
->
Service
to access CSD.
4.) Log in as GE Service, enter current password
5.) Select
Diagnostics
tab.
6.) Select
Service
->
Diagnostics
.
7.) Select the
“+”
before
Service Diagnostics
folder
to open the list of choices.
8.) Select the
“+”
before
Analog Receive
to open the test choices for the analog receiver.
9.) Select
DC Offset Calibration
and then follow the on-screen instructions.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......