Section 8-9 - BEP (Back End Processor) parts replacement
8-9-2-2
BEP installation - R3.x and earlier
Table 8-123 BEP installation - R3.x and earlier
Steps
Corresponding Graphic
1.
Reinstall the following parts from the BEP that was replaced (all that were present):
• DVR Card
• Flex Jumper Cable
• SATA cable (which goes between the DVR and Motherboard)
• Patient I/O filler
• Patient I/O module
• Option Dongle
• Printer Tray
2.
Slide the BEP into the left side of the chassis frame.
3.
Install the two HEX key screws at the inside
base of the BEP (Torque: 10 Nm
{7.4 lbf-ft}).