Section 10-6 - Electrical Safety Tests
10-6-3 Leakage Current Limits (cont’d)
NOTE:
Leakage current limits for Ultrasound system operation on 230-240 Volt mains
*ISO (on Dale 600) and Mains Applied (on Dale 601) refer to the sink leakage test where mains (supply)
voltage is applied to the part to determine the amount of current that will pass (or sink) to ground if a
patient contacted mains voltage.
Table 10-11 ISO (on Dale 600) and Mains Applied (on Dale 601) Limits*
Probe Type
Measurement
BF
5.0 mA (5000 µA)
CF
0.05 mA (50 µA)