GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 2 Site preparations
2 - 1
Chapter 2
Site preparations
Section 2-1
Overview
2-1-1
Purpose of this chapter
This chapter provides the information required to plan and prepare for the setup of a LOGIQ E9.
Included are descriptions of the facility and electrical needs to be met by the purchaser of the LOGIQ E9.
Section 2-2
General console requirements
2-2-1
Console environmental requirements
2-2-1-1
If the LOGIQ E9 is very cold or hot
When unpacking the device, allow the temperature of the device to stabilize before powering up. The
following table describes guidelines for reaching operational temperatures from storage or transport
temperatures. See:
Table 2-1
’
2-2-1-2
Environmental specifications
Table 2-1
LOGIQ E9 Acclimate Time
o
C
60
55
50
45
40
35
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
-35
-40
o
F
140
131
122
113
104
95
86
77
68
59
50
41
32
23
14
5
-4
-13
-22
-31
-40
Hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20
Table 2-2
Environmental Specifications for LOGIQ E9 Scanners
Operational
Storage
Transport (< 16 hrs.)
Temperature
10º - 35º C
(50º - 96º F)
-10º - 50º C
(-14º - 122º F)
-10º - 50º C
(-14º - 122º F)
Humidity
30% - 80%
non-condensing
30% - 80%
non-condensing
30% - 80%
non-condensing
Pressure
70 - 106 kPa
70 - 106 kPa
70 - 106 kPa
Heat Dissipation
4712 Btu/hour
“
“
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......