Section 3-6 - Completing the setup
3-6-4
Connections on the I/O Rear Panel
NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according
to the respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-
1 for medical equipment). Furthermore, all complete configurations shall comply with the valid
version of the system standard IEC60601-1-1. Everybody who connects additional equipment
to the signal input part or signal output part of LOGIQ E9, configures a medical system, and is
therefore responsible that the system complies with the requirements of the valid version of
IEC60601-1-1. If in doubt, consult the technical service department or your local representative
for GE Healthcare.
3-6-4-1
Connect Ethernet
Connect the Ethernet cable to the Ethernet connector (
3
) on the External I/O (rear side of LOGIQ E9).
Figure 3-7 Ethernet, Audio (1) and DVI-I (2) connection for External Monitor on
rear side of LOGIQ E9 - R3.x and earlier
!
R
L
5180173 revx
1
3
2