Section 2-2 - General console requirements
2-2-4
Probes environmental requirements
2-2-5
Time and manpower requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
Table 2-6
Environmental Requirements - Probes
Standard Probes
4D Probes
Operation:
10
o
to 40
o
C (50 to 104 ºF)
18
o
to 40
o
C (64.4 to 104 ºF)
Storage:
-10° to 60° C (14 to 140 ºF)
-10° to 60° C (14 to 140 ºF)
-10
o
to 50
o
C (14 to 122 ºF)
Temperatures in degrees Celsius (
o
C) conversion to degrees F: (
o
F) = (
o
C * 9/5) + 32
NOTICE
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO +50 degrees C. or +60 degrees C, DEPENDING ON THE TYPE OF PROBE. WHEN
EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT IN
ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
Refer to the Table in
section 2-2-1-1 on page 2-1
to determine the needed settlement time.
CAUTION
HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE LOGIQ E9.
ATTEMPTS TO MOVE THE LOGIQ E9 CONSIDERABLE DISTANCES OR ON AN INCLINE BY ONE
PERSON COULD RESULT IN INJURY OR DAMAGE OR BOTH.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......