GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 1 Introduction
1 - 19
1-4-4
Electrical safety
1-4-4-1
Safe practices
Follow these guidelines to minimize shock hazards whenever you are using the scanner:
•
The equipment chassis must be connected to an electrical ground.
•
The unit is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. A separate power outlet with a 20 amp circuit breaker for 120
VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or 15 amp circuit
breaker for 100 VAC for Japan.
•
The power outlet used for this equipment should not be shared with other types of equipment.
•
Both the system power cable and the power connector must meet international electrical standards.
1-4-4-2
Probes
Follow these guidelines before connecting a probe to the scanner:
•
Inspect the probe prior to each use for damage or degradation to the:
-
housing
-
cable strain relief
-
lens
-
seal
-
connector pins
-
locking mechanism
•
Do not use a damaged or defective probe.
•
Never immerse the probe connector or adapter into any liquid.
•
The system has more than one type of probe port. Use the appropriate probe port designed for the
probe you are connecting.
WARNING
WARNING
Connecting a LOGIQ E9 scanner to the wrong voltage level will most likely destroy it.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......