Removable Fan Tray Filter cleaning
Clean the Fan Tray filter to ensure that a clogged filter does not cause the system to overheat and
reduce system performance and reliability. It is recommended the filter be cleaned quarterly (once every
three months) or more often in locations where high levels of dust are present.
Table 8-112 Removable Fan Tray Filter cleaning
Manpower /
Time
Tools
One person /
15 minutes
No tools are needed to replace the filter.
Table 8-113 Preparations and Preparation Link
Preparations - you must perform the following steps
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet.
3. Remove the Removable Fan Tray.
Preparation Link (if you need more information):
4-2-3 "Power shut down" on page 4-7.
Table 8-114 Removing the Fan Tray Filter
Steps
Corresponding Graphic
1.
Slide Fan Tray out of the LOGIQ E9.