GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 4 Functional Checks
4 - 35
4-3-6
M-Mode Checks
For information on the system’s different modes as well as expected results, the Basic User Manual or
User Guide will familiarize you with image optimization for
B-Mode
,
M-Mode
,
Color Flow
, and
Doppler
.
4-3-6-1
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
-
see:
Section 3-6-6 "Connecting Probes" on page 3-20
for info about connecting the probes.
-
For available probes, see: On-line Help or Chapter !7 of the Basic User Manual.
2.) Turn ON the scanner.
4-3-6-2
Checks
•
Check Horizontal Sweep, Frequency, and Focus
•
Check Compress, Reject, Power and Dynamic Range
Table 4-15 B/M-Mode Functions
Control
Description
Power Output (Acoustic Power)
Optimizes image quality and allows user to reduce beam intensity. 10% increments between 0-100%. Values
greater than 0.1 are displayed
Dynamic Range
Controls how echo intensities are converted to shades of gray, thereby increasing the adjustable range of contrast.
Focus Number and Position
Increases the number of transmit focal zones or moves the focal zone(s) so that you can tighten up the beam for a
specific area. A graphic caret corresponding to the focal zone position(s) appears on the right edge of the image.
Rejection
Selects a level below which echoes will not be amplified (an echo must have a certain minimum amplitude before
it will be processed).
Frame Average
Temporal filter that averages frames together. This has the effect of presenting a smoother, softer image.
Colorize
Enables gray scale image colorization. To deactivate, reselect a Gray Map.
Gray Map
Determines how the echo intensity levels received are presented as shades of gray.
Rotation (Up/Down)
Rotates the image by selecting the value from the pop up menu.
Frequency
Multi Frequency mode lets you downshift to probe's next lower frequency or shift up to a higher frequency.
Frame Rate/Resolution
Optimizes B-Mode frame rate or spatial resolution for the best possible image.
Sweep Speed
Changes the speed at which the timeline is swept.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......