GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 205
4.
Reach inside the BEP and connect:
• the BEP cable to the Backplane
J3
on a
GFI Configuration.
• the BEP cable to Backplane
J2
(left) on a
MRX Configuration.
BEP Cable to Backplane J3 of GFI Configuration
(views inside the BEP)
BEP Cable to Backplane, J2 (left) of
MRX Configuration
5.
NOTE:
If you DO NOT reconnect the internal BEP cable to the Backplane, the system will power
up but WILL NOT scan or WILL LAUNCH into simulator mode.
6.
Insert the bottom lip of the BEP cover inside the base of the BEP.
7.
Connect the Battery cable to
J3
on the EPS and EPS to BEP cables to the EPS.
8.
MAKE SURE all cables are clear of the BEP Cover and tilt the top of the BEP cover toward the
BEP.
NOTICE
Be careful not to pinch any of the cables when installing the BEP cover.
9.
Tighten the thumb screw or screws at the top of the BEP cover that attach the cover to the BEP.
Table 8-123 BEP installation - R3.x and earlier
Steps
Corresponding Graphic